![background image](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985210.webp)
5-58
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Set Up Instructions
Section 5
Interpretive Report Messages
If the
PRINT INTERPRETIVE REPORT
option on the
CUSTOMIZE PRINTED
REPORT
screen for the Graphics Printer is enabled, Interpretive Report messages
are printed on the report. Some of these messages are generated from the
information entered for the Patient Limits. The following messages are generated
by specimen results which exceed the Patient Limits:
WBC Messages
Leukopenia —
result exceeds the lower limit for WBC
Leukocytosis —
result exceeds the upper limit for WBC
Neutropenia —
result exceeds the lower limit for Neutrophil absolute number
Neutrophilia —
result exceeds the upper limit for Neutrophil absolute number
Lymphopenia —
result exceeds the lower limit for Lymphocyte absolute number
Lymphocytosis —
result exceeds the upper limit for Lymphocyte absolute number
Monocytosis —
result exceeds the upper limit for Monocyte absolute number
Eosinophilia —
result exceeds the upper limit for Eosinophil absolute number
Basophilia —
result exceeds the upper limit for Basophil absolute number
RBC Messages
Anemia —
result exceeds the lower limit for RBCs
Polycythemia —
result exceeds the upper limit for RBCs
Microcytic RBC —
result exceeds the lower limit for MCV
Macrocytic RBC —
result exceeds the upper limit for MCV
Hypochromic —
result exceeds the lower limit for MCHC
Hyperchromic —
result exceeds the upper limit for MCHC
Anisocytosis —
result exceeds the upper limit for RDW
PLT Messages
Thrombocytopenia
—result exceeds the lower limit for PLTs
Thrombocytosis
—result exceeds the upper limit for PLTs
Microcytic PLT
—result exceeds the lower limit for MPV
Macrocytic PLT
—result exceeds the upper limit for MPV