10-6
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Diagnostics Menu
Section 10
Operator correctable faults (for example, waste full, diluent empty) can be cleared
by pressing the
[CLEAR FAULT]
key after taking corrective action. After corrective
action for a fatal fault, the system must be reinitialized by pressing the
[INITIALIZATION]
soft key in the second
DIAGNOSTICS MENU
screen (refer to
Execution Times
This key is for service use only.
Figure 10.5 Count Rate Summary Screen
CNT Rate Summary
When the
[CNT RATE SUMMARY]
key is pressed, the following soft key labels (see
Figure 10.5) are displayed:
WOC CNT RATE/WOC CNT GRAPH/WOC CNT DATA *
RBC/PLT CNT RATE/RBC/PLT CNT GRAPH/RBC/PLT CNT DATA *
NOC CNT RATE/NOC CNT GRAPH/NOC CNT DATA *
PRINT
RETURN
* Key labels alternate between the selections.
NOTE:
Once a
[CNT RATE]
key is pressed, that key toggles between
[CNT DATA]
and
[CNT GRAPH]
until the operator exits the screen. The keys then revert
to
[CNT RATE]
.