CELL-DYN
®
3200 System Operator’s Manual
6-33
9140181H—October 2001
Section 6
Calibration Procedures
Enter Factor Method
Overview
The Enter Factor Calibration Method is used to enter a predetermined factor to
adjust calibration when a consistent bias exists between the CELL-DYN 3200
System and a comparison analyzer. This method can also be used when calibrating
the secondary mode of the system to the primary mode.
Either Calibrator or Fresh Whole Blood can be used for calibration.
Fresh Whole Blood: Because the same samples will be used to first obtain
reference values on a reference instrument then to calibrate in the Open and Closed
Modes, it is important to begin with a sufficient amount of sample (12 mL each is
recommended).
A set of worksheets is provided in
Manual Calibration Worksheets
at the end of
this section to assist in the calibration process.
NOTE:
Always complete the Pre-Calibration procedures before beginning any
calibration.
Enter Calibration Factor Screen
When the Enter Factor key is pressed in the main
CALIBRATION
menu, the
ENTER
CALIBRATION FACTO
R menu is displayed (refer to Figure 6.11). New calibration
factors are manually entered on this screen.
The following soft key labels are displayed:
RESTORE FACTORS
RESET ALL TO 1.000
RETURN