![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 200](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985200.webp)
5-48
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Set Up Instructions
Section 5
Figure 5.22 Customize Printed Report Screen for the Graphics Printer
Graphics Printer
When the
[GRAPHICS PRINTER]
key is pressed, the
CUSTOMIZE PRINTED REPORT
screen for the Graphics Printer is displayed. (See Figure 5.22.) A description of the
options available on the
CUSTOMIZE PRINTED REPORT
screen
is given below:
1.
AUTO-PRINT RESULTS FOR ALERTED SPECIMENS
When this option is enabled, a report is automatically printed for any sample
with flagged results.
2.
AUTO-PRINT RESULTS FOR NON-ALERTED SPECIMENS
When this option is enabled, a report is automatically printed for any sample
without flagged results.
3.
PRINT GRAPHS FOR ALERTED SPECIMENS ONLY
When this option is enabled, scatterplots and histograms are printed only for
samples with flagged results.
4.
PRINT PCT, PDW RESULTS
When this option is enabled, the results for PCT and PDW are printed on the
report.
NOTE:
Clinical significance has not been established for these parameters.
Therefore, they are not reportable.
5.
PRINT X-B RBC PROGRAM STATUS
When this option is enabled, the status of the X-B RBC program is printed on
the report. The X-B status (for example, X-B RBC: 13/OUT2) is printed
below the Status Box.