10-34
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Table 10.1 Status
Conditions
Each message is followed by one of the abbreviations shown below indicating
where that message appears on the screen:
BL — Bulletin Line
SB — Status Box
Ready
/SB
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The Analyzer is ready to process samples.
Selecting open/closed mode
/SB
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The
[CHANGE SAMPLER]
key was pressed and the Analyzer is changing to the
selected mode of operation.
Resume processing when the READY message is displayed in the Status Box.
Entering standby
/SB
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The instrument has been idle for four hours and therefore is automatically
performing a cleaning cycle before entering the STANDBY state. (Pressing the
[DAILY SHUTDOWN]
key also initiates this message.)
When the cycle is finished, press
[RUN]
to prime the system and bring the
instrument to the READY state. Resume operation when the cycle is finished.
Standby
/SB
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The instrument has entered the STANDBY state.
Press
[RUN]
or
[PRIME]
to prime the system and bring the instrument to the
READY state. Resume operation when the cycle is finished.
Initialized
/SB
The Analyzer hardware initialization is completed.
Press
[RUN]
to prime the system and bring the instrument to the READY state.
Resume operation when the cycle is finished.
Unpinching valves
/SB
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The Analyzer was in standby for a predetermined time period and therefore the
valves are being exercised to be sure that tubing is not pinched shut.
Analyzer status will change back to STANDBY when the unpinching operation
is completed.