![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 455](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985455.webp)
CELL-DYN
®
3200 System Operator’s Manual
10-21
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Stop Transmiss
The
[STOP TRANSMISS]
key is used to stop any transmission that is in progress to
a LIS. When the key is pressed, any transmission in progress is aborted.
Figure 10.20 Serial Test Transmit Message Screen
Transmit Message
When the
[TRANSMIT MESSAGE]
key is pressed, the message
“CELL-DYN serial interface test”
is transmitted from the Analyzer
to the COM1 port, through the loop-back connector and back to the Analyzer. The
screen (see Figure 10.20) displays the message:
MESSAGE SENT: CELL-DYN serial interface test.
If the test is successful, the screen displays the message:
MESSAGE RECEIVED: CELL-DYN serial interface test.
This message indicates that the Data Station is communicating properly. If the test
is not successful, no message appears.