5-114
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Using The Work List
Section 5
8.
DEFAULT PATIENT LIMIT SET (1 to 6)
This field is used to specify the default (pre-assigned) Patient Limit Set that
is automatically assigned to each sample unless otherwise indicated in the
Work List. Any number between 1 and 6 inclusively may be entered.
9.
PARAMETER SET ENTRY SELECTED
If this option is turned OFF, the prior setting will be entered automatically for
each entry. The operator can, however, input a different setting for individual
entries by using the arrow keys to place the cursor in this field and type in a
new setting. The new setting applies only to the entry(ies) that was changed.
If this option is turned ON, the operator may then select another default
setting by typing in a number from 2 to 4. This new setting will be
automatically assigned to each new entry. The operator also has the ability to
enter a different setting for individual entries. When the cursor moves to the
<Parameter Set> field of a particular entry, the operator may override the
default setting and type in a new setting for that entry.
10. DEFAULT PARAMETER SET (1 to 4)
This field is used to specify the default (pre-assigned) Parameter Set that is
automatically assigned to each sample unless otherwise indicated in the Work
List. Any number between 1 and 4 inclusively may be entered.