6-16
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Calibration Procedures
Calibration Menu
Section 6
Calibration Log
The
[CALIBRATION LOG]
key is used to display the
CALIBRATION LOG
screen for the
Open or Closed Mode (refer to Figure 6.4). The soft keys available in this screen
are:
OPEN SAMPLER/CLOSED SAMPLER
PRINT LOG
RETURN
The
CALIBRATION LOG
holds 10 entries. The last 5 entries are displayed on the
screen. To display any previous entries, press the Page Up key on the keyboard.
NOTE:
When the log is full, subsequent entries cause the oldest entry to be
deleted and the remaining entries to move up one line, so that the current
factors are added to the bottom of the list. Therefore, the log should be
printed periodically for purposes of documentation.
The log displays the
DATE
,
TIME
,
OPERATOR ID
,
CALIBRATION FACTORS
and a line for
COMMENTS
.
NOTE:
The letters in parentheses after each factor indicate the method of factor
derivation: F = Factory, A = Auto-Cal, E = Enter Factor
(manual factor entry).
Type any comments in the
COMMENTS
field. Press the Enter key on the keyboard
to save the entry and advance the cursor.
NOTE:
Comments may be added to the Calibration Log only after a calibration
factor is changed or reentered.
Open Sampler/Closed Sampler
The
[OPEN SAMPLER/CLOSED SAMPLER]
key is used to display the calibration
factors for the displayed mode. This key toggles between the Open and Closed
Sampler Calibration Factors.
Print Log
The
[PRINT LOG]
key is used to print the Calibration Log for the displayed mode.