![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 123](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985123.webp)
CELL-DYN
®
3200 System Operator’s Manual
3-29
9140181H—October 2001
Section 3
Principles of Operation
Parameter Flagging Messages
Table 3.3 summarizes all of the parameter flagging messages by parameter and
category.
Table 3.3
Parameter Flagging Messages
Parameter
Dispersional Data Alerts
Suspect
Parameter Flags
Suspect
Population Flags
Interpretive
Messages
WBC
Result displays in yellow if
below lower limit
Result displays in purple if
above upper limit
Result underlined on graphics
printout when limits exceeded
Result underlined on blank
ticket when limits exceeded
Result marked with asterisk (*)
on preprinted ticket when
results exceeded
WBC
NWBC
FWBC
NRBC
RRBC
Leukopenia
Leukocytosis
Differential
NEU
LYM
MONO
EOS
BASO
Same as WBC
DFLT (NLMEB)
DFLT (NE)
DFLT (LM)
DFLT (B)
DFLT (LB)
BAND
IG
BLAST
VAR LYM
Neutropenia
Neutrophilia
Lymphopenia
Lymphocytosis
Monocytosis
Eosinophilia
Basophilia
RBC
HGB
MCV
RDW
Same as WBC
RBC MORPH
Anemia
Polycythemia
Microcytic RBC
Macrocytic RBC
Hypochromic
Hyperchromic
Anisocytosis
PLT
MPV
Same as WBC
LRI
URI
LURI
The MPV value
may be suppressed
(not displayed
or printed).
Thrombocytopenia
Thrombocytosis
Microcytic PLT
Macrocytic PLT