CELL-DYN
®
3200 System Operator’s Manual
10-41
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Uninitialized/
SB
– – – – Probable Cause – – – –
– – – – Corrective Action – – –
1. Main power switch is ON but the
Analyzer is not responding.
1. Turn the system OFF.
2. Ensure that all cables on the back of
the instrument are properly
attached.
3. Turn the System ON.
4. If the problem persists, contact your
Abbott Hematology Customer
Support Center.
Initialization Failed
Bottom of screen (MAIN MENU is not displayed)
– – – – Probable Cause – – – –
– – – – Corrective Action – – –
1. The instrument software was unable
to initialize. The CELL-DYN
software does not display the
MAIN
MENU
screen.
1. Initialize the Analyzer by following
the instructions in the Power OFF
and Power ON procedures given in
Subsection:
Operation Overview
.
If the Analyzer does not initialize,
press the Print Screen key on the
computer keyboard to document
any screen messages, then contact
your Abbott Hematology Customer
Support Center.
Table 10.2 General Fault Conditions (Continued)