Tab 2
2.5 List of error messages
2.5
E7
61
12
Unit is not ready for operation
This error is a sequential fault.
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON.
z
Repeat procedure and observe causal error
messages.
z
Checking the CAN bus
z
Replace diaphragm unit
E5
61
15
Drive overtemperature
Note:
Can occur after frequent diaphragm
travel, e.g. during demo operation at a
fair
z
Allow unit to cool down.
If the error occurs again...
z
Check board DX61, replace if necessary
Replace board DX61 or
Replace diaphragm unit
E5
61
20
Horizontal motor malfunction during
operation
Note:
Can occur after frequent diaphragm
travel, e.g. during demo operation at a
fair
z
Allow unit to cool down.
If the error occurs again...
z
Check board DX61, replace if necessary
Replace board DX61 or
Replace diaphragm unit
E5
61
21
Horizontal motor limit switch error
z
Check light barrier V61_1 and connectors, re-
place if necessary
z
Perform diaphragm adjustment
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
If the error occurs again...
z
Check board DX61, replace if necessary
Replace board DX61 or
Replace diaphragm unit
E5
61
22
Horizontal motor movement error
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Check board DX61, replace if necessary
Replace board DX61 or
Replace diaphragm unit
E5
61
30
Vertical motor malfunction during
operation
z
Allow unit to cool down.
If the error occurs again...
z
Check board DX61, replace if necessary
Replace board DX61 or
Replace diaphragm unit
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...