7.5 Checking the light localizers
Tabs 7
7.5
Checking the light localizers
Check whether the FH light localizer
(PAN) can be adjusted easily
z
Adjust the FH light localizer on the PAN with slider
A
.
If the light localizer cannot be moved easily:
Æ
Lubricate the slider (rail) with a light coat of Vaseline
FH light localizer PAN
MS light localizer
Light localizer Ceph
PAN
A = FH light
MS light
localizer
localizer