2.5 List of error messages
Tab 2
E5
13
36
Software signal of key is applied, but
hardware signal is not
z
Check cables L9 and L10,
replace if necessary
z
Check limit switches SE1_1 and SE1_2,
replace if necessary
z
Replace the user interface
(Easypad for XG
Plus
,
Multipad for XG5 / XG 3)
E7
13
37
Overtravel of HA motor occurs or
height adjustment power transistor
defective
z
Check HA motor for overtravel,
replace board DX1 if necessary
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON and check functioning.
z
Replace board DX1
E6
13
38
Height adjustment motor is not ready
for operation
This error is a sequential fault.
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON.
z
Repeat procedure and observe causal error
messages.
E6
13
39
Error when activating height adjust-
ment motor
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON and check functioning.
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Replace board DX1
E3
13
40
Release signal applied during power-
on.
z
Unit restart:
Switch the unit OFF. Wait for 30 sec.
Switch unit ON, making sure that the release
button is not pressed during boot-up.
If the error occurs again...
z
Check the X-ray signal path
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...