båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 35
Tab 2
2.5 List of error messages
2.5
E6
41
07
TTP* timeout error
(detected on slave side)
Note:
The module was temporarily not
addressed by the master:
- Undervoltage on the master side
- Procedure error in the software
- Master (DX11) receives no
return commanding from the module
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Check power supply of board DX11; measuring
point 3.3 V on board DX1 (see wiring diagrams).
– If 3.3 V are present, replace board DX11
– If 3.3 V are not present, replace board DX1
E6
41
08
General fault detected locally on mod-
ule (slave side). CAN controller being
reinitialized.
z
Checking the CAN bus
z
Check software versions via info screen or ser-
vice routine S008.2,
perform a software update if necessary
z
Replace board DX41
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
41
09
Board DX41 is installed in the unit, but
not configured
z
Perform device configuration via service routine
S017.9
E7
41
10
Module is stuck in bootloader stage
z
Check operating status of board
(note LED states)
If the board remains in the bootloader stage...
z
Repeat software update
z
Replace board DX41
E7
41
12
Unit is not ready for operation
This error is a sequential fault.
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON.
z
Check unit configuration (with or without DX41)
via service routine S017.9,
configure correctly if necessary
If the error occurs again...
z
Repeat procedure and observe causal error
messages.
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...