5.14 Service routine S018
5.14
S018.2: Set minimum travel height
5.14.4
S018: Test step 5
Moving the unit
1.
Move the unit to the required minimum travel height by pressing the
Up/Down keys
in the user mode.
NOTE
i
Programming the minimum travel height is possible only for a system height that is
below the lower correction switch level (< position value of 1500)!
Selecting test step 5
2.
Select test step 2 using the
arrow keys
in selection field 5
and then confirm your selection with the
Service key
.
The current height position is displayed in selection field 1. The
Memory
key
(XG
Plus
) or the LED above the
Memory key
(XG 5 / 3 / 3 PPE)
lights up.
T
R
Prog.
S
kV
mA
(
n
(
n
1
2
3
+
-
S018
5
n
(
1
2
3
S018_______5____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
+
-
(
n
(
n
(
o
(
o
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...