59 38 399 D3352
2 – 50
D3352.076.01.13.02
07.2008
2.5 List of error messages
Tab 2
Location 81, zz = 1-23: Board DX81, pan slot
Error code
Description
Actions required
see page
E6
81
01
PAN
General error during module initializa-
tion
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Check pan sensor, replace if necessary
E6
81
02
PAN
Invalid system data or uninitialized
module storage data
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Check pan sensor, replace if necessary
E6
81
03
PAN
Invalid commanding or control data
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
If the error is a software error known to the Siro-
na Customer Service Center (CSC), a software
update (bugfix) must be performed
E6
81
04
PAN
Data transfer error or dialog error to
module (master side)
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
81
05
PAN
Data transfer error or dialog error to
bootloader of module
Note:
Occurs only in connection with
software update
z
Repeat software update
z
Checking the CAN bus
z
Check pan sensor, replace if necessary
E6
81
06
PAN
Module failed in TTP*
(detected on master side)
z
Checking the CAN bus
z
Check pan sensor, replace if necessary
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...