59 38 399 D3352
3 – 24
D3352.076.01.13.02
07.2008
3.9 Error analysis of X-RAY control signal path: unit HW version BA or higher
3.9.3
Error messages
with and without installed
remote control
Error code
Description
Actions required
see page
E3
13
40
Short circuit in signal path between
release button A2 and board DX11
during switch-ON.
z
Replace cable L117 or L108
z
Replace board DX1
z
Replace board DX11
E6
13
41
Release signal missing on board
DX11 at start of exposure.
z
Replace cable L117 or L108
z
Replace board DX1
z
Replace board DX11
E3
13
42
Short circuit in signal path between
release button A2 and board DX11
during unit operation.
z
Replace cable L117 or L108
z
Replace board DX1
z
Replace board DX11
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...