båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 53
Tab 5
5.8 Service routine S008
5.
8
S008.3: Input/confirmation of unit serial number
6.
Check the correctness of the unit serial number you have entered and, if it
is OK, confirm this by pressing the
R key
.
CAUTION
Make sure that you do not accidentally press the R key and/or the Memory key
before all digits of the serial number have been entered. Otherwise an
incorrect serial number will be irreversibly entered.
Case B:
Confirming the unit serial number
1.
To confirm the displayed serial number, press the
R key
.
You can interrupt this procedure with the Service key
.
The unit serial number will then not be confirmed.
Quitting the service routine
XG
Plus
:
To return to the service routine selection menu, press the
Service key
or
the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
For XG 5 / 3 / 3 PPE:
Pressing the
up arrow key
above selection field 3 returns you to the ser-
vice routine selection menu.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
R
R
1
2
3
000000123
n
(
1
2
3
000000123_______
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
R
3
(
n
R
(
n
)
n
S8
R
R
1
2
3
000000123
n
(
1
2
3
000000123_______
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
R
3
(
n
R
(
o
)
o
S8
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...