båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 25
Tab 2
2.5 List of error messages
2.5
E5
13
31
Unit has traveled to upper limit switch
z
Check max. travel height with service routine
S018.2, adjust if necessary
z
Run HA motor in the other direction with the UP/
DOWN keys and reference (value approx.
1500)
z
Check light barriers V1_4
replace if necessary
z
Check HA motor for overtravel,
replace DX1 if necessary
If the error occurs again...
z
check the limit switch or wiring,
correct or replace the limit switch if necessary
E5
13
32
Unit has traveled to lower limit switch
z
Run HA motor in the other direction with the UP/
DOWN keys and reference (value approx.
1500)
z
Check light barriers V1_4
replace if necessary
z
Check HA motor for overtravel,
replace DX1 if necessary
If the error occurs again...
z
check the limit switch or wiring,
correct or replace the limit switch if necessary
E5
13
33
Height adjustment motor position
counter too small for current position
Note:
Error may occur after replacement of
board (DX11)
z
Run HA motor in the other direction with the UP/
DOWN keys and reference (value approx.
1500)
z
Check max. travel height using service routine
S018.2, adjust if necessary
z
Check light barriers V1_4,
replace if necessary
E5
13
34
Height adjustment motor position
counter too large for current position
Note:
Error may occur after replacement of
board (DX11)
z
Run HA motor in the other direction with the UP/
DOWN keys on the Easypad (XG
Plus
) or the
Multipad (XG5/3/3 PPE) and reference (value
approx. 1500)
z
Check light barriers V1_4,
replace if necessary
HINWEIS
i
As an aid, the current switching state can be queried
via service routine S018.4.
E5
13
35
Height adjustment motor;
wrong direction of rotation
z
Check connector assignment on filter or in front
of HA motor,
correct if necessary
z
Replace board DX1
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...