59 38 399 D3352
2 – 36
D3352.076.01.13.02
07.2008
2.5 List of error messages
Tab 2
*)
TTP = Time Trigger Protocol
E3
41
20
Release signal applied during power-
on
z
Unit restart:
Switch the unit OFF. Wait for 30 sec.
Switch unit ON, making sure that the release
button is not pressed during boot-up.
If the error occurs again...
z
Check signal path for short-circuit according to
wiring diagrams (see section "Error analysis of
X-RAY control signal path: up to unit hardware
revision AG (with board DX41)"),
replace component if necessary
E6
41
21
CAN bus connection to board DX41
interrupted.
Board DX41 cannot address board
DX42 via the separate CAN bus con-
nection.
z
Check cable L17,
replace if necessary
z
Check board DX42,
replace if necessary
z
Check board DX41,
replace if necessary
E6
41
23
Hardware fault at controller input on
board DX41.
Board DX41 detects a wrong signal
level of the hardware signal for radia-
tion release.
z
See section
"Error analysis of X-RAY control
signal path: up to unit hardware revision AG
(with board DX41)"
E6
41
24
Short circuit in radiation release signal
path between board DX42 and board
DX41 (cable L17).
The release signal was detected on
boards DX11 and DX41 but not on
board DX42.
z
See section
"Error analysis of X-RAY control
signal path: up to unit hardware revision AG
(with board DX41)"
E6
41
25
X-Ray hardware signal present,
software signal not present
z
Check cable L17,
replace if necessary
z
Check board DX42,
replace if necessary
z
Check board DX41,
replace if necessary
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...