59 38 399 D3352
4 – 26
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
Coarse or precision adjustment?
In most cases, the diaphragm can be adjusted using precision adjustment
from the start (see step 5 onward). Only in exceptional cases, e.g. if the
exposed image area is completely outside the image field
C
in an image
acquired with
PRECISION
ADJUSTMENT
setting, is it necessary to perform a
coarse adjustment prior to precision adjustment
D
. To do this, deactivate the
PRECISION
ADJUSTMENT
check box (see page 4-3) and then perform a
coarse adjustment proceeding in the same way as for precision adjustment.
The only difference between coarse
and precision adjustment is the size of
the image area considered. Furthermore, there are fewer auxiliary lines in the
coarse adjustment mode.
X-ray image with unadjusted pan diaphragm:
Alternatively with precision adjustment setting (C) and coarse adjustment setting (D)
On the X-ray image with coarse adjustment D, the exposed area is just barely
visible at the right margin of the image field. Even in this extreme case, an
adjustment would still be possible.
NOTE
i
With SIDEXIS SW V01.45 and higher, a message window indicates whether a
coarse adjustment is required on completion of the precision exposure.
NOTE
i
When the
diaphragm is correctly adjust-
ed
, a course adjustment does not produce
a meaningful image, since no surrounding
border can be seen in this case (similar to
image C).
C
D
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...