D3352.076.01.13.02
07.2008
4 – 51
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
NOTE
i
If the X-ray image still does not correspond to the ideal image, it is pos-
sible that one or more criteria have not yet been fulfilled:
Case 1: Center needle not coincident with auxiliary line
If the shadow of the center needle, the needle image and the auxiliary line do
not yet coincide, then repeat the adjustment procedure starting with step 5.
Case 2: Symmetry not OK
If the shadow of the center needle, the needle image and the auxiliary line co-
incide, but the symmetry or the distance between the two outer needles is not
yet correct, then repeat the adjustment procedure starting with step 5.
NOTE
i
If you do not reach your goal via automatic adjustment, repeat the adjustment
procedure with manually determined adjustment values (see page 4-52).
Saving the values
8.
If all criteria are fulfilled and the current image is identical to the ideal im-
age
A + B
, then save the values:
Click
SAVE
VALUES
NOTE
i
In SIDEXIS versions V01.50 and higher, the values for
S1
,
S2
and
S3
in the
PAN
-
SYMMETRY
submenu are set
equal to zero
with a correct adjustment,
i.e. if they are within the permissible tolerance. For versions < V01.50, the
values in the text boxes may deviate slightly from zero even with a correct
adjustment.
z
Go on to the next adjustment step.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...