båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 15
Tab 2
2.5 List of error messages
2.5
*)
TTP = Time Trigger Protocol
E7
07
21
No CAN bus connection.
DX11 does not start.
Note:
Occurs after power-on in the start
screen.
z
Start the detail query via Sixabcon
If DX11 responds...
z
Check signal path to DX7,
repair or replace cable/connector if necessary
z
Replace DX1
If DX11 does not respond...
z
Replace DX11
E3
07
30
Height adjustment keys actuated dur-
ing power-on
z
Unit restart:
Switch the unit OFF. Wait for 30 sec.
Switch unit ON, making sure that the Easypad
is not actuated during boot-up.
If the error occurs again...
z
Check user interface with electronics (DX7) or
Replace Easypad
E3
07
31
Forehead support key pressed during
power-on
E3
07
32
Temple support keys actuated during
power-on
E3
07
33
Light localizer key actuated during
power-on
E3
07
34
Test key actuated during power-on
E3
07
35
Return key actuated during power-on
E3
07
36
Touchscreen actuated during power-
on
E6
07
40
No valid language set found
z
Check selected language set using service rou-
tine S017.5, correct if necessary
HINWEIS
i
Check whether selected language set is already in-
stalled, perform software update if necessary.
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...