2.5
z
Check complete mechanics/spindle of motor
M91_3 for smooth and easy running,
replace if necessary
z
Replace board DX91
E5
91
16
Motor control timeout (M91_2/3)
z
Replace board DX91
E5
91
20
Sensor drive malfunction during oper-
ation
z
Allow unit to cool down.
If the error occurs again...
z
Check complete mechanics/spindle of motor
M91_3 for smooth and easy running,
replace if necessary
z
Check light barrier V91_2,
E5
91
21
Sensor drive, opposite light barrier
reached prematurely
replace if necessary
z
Check cable to light barrier (L21/L24),
replace if necessary
z
Check board DX91,
replace if necessary
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E5
91
22
Sensor drive timeout
z
Check motor M91_3 incl. cable and connector,
replace if necessary
z
Check cable to light barrier (L21/L24),
replace if necessary
E5
91
30
Secondary diaphragm drive malfunc-
tion during operation
z
Allow unit to cool down.
If the error occurs again...
z
Check motor M91_2 incl. cable and connector,
replace if necessary
z
Check cable to light barrier (L21/L24),
replace if necessary
E5
91
31
Secondary diaphragm drive, opposite
light barrier reached prematurely
z
Check light barrier V91_2,
replace if necessary
z
Check cable to light barrier (L21/L24),
replace if necessary
z
Check board DX91,
replace if necessary
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...