59 38 399 D3352
3 – 18
D3352.076.01.13.02
07.2008
3.8 Error analysis of X-RAY control signal path: up to unit hardware revision AG
3.8
Error analysis of X-RAY control signal path:
up to unit hardware revision AG
(with board DX41)
3.8.1
Error and help messages
with remote control
installed
E3 42 31 + E3 13 40 + E3 41 20
occur in combination after the unit
is switched ON with the door
contact closed.
E3 42 31
occurs individually after the unit is
switched ON.
yes
no
Was the release but-
ton actuated during
switch ON?
Switch the unit OFF
and then ON again.
Make sure that the
release button is not
actuated during
switch-ON.
Do the error messages
recur?
yes
Short circuit in the coiled cable of the release
button or on the membrane keyboard of the
remote control:
• Replace release button
• Replace remote control
Hardware fault on display board DX42 or short circuit in coiled
cable of release button or on membrane keyboard of the remote
control:
• Replace release button (unit hardware version
<
BA)
• Replace remote control (see section 6.20)
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...