59 38 399 D3352
5 – 138
D3352.076.01.13.02
07.2008
5.19 Service routine S034
Tabs 5
S034.4: Display/calibrate center position of ceph scan sensor axis
Checking the new center position of the
ceph scan (sensor axis)
9.
Quit the test mode by pressing the
T key
.
NOTE
i
The offset settings are taken into account in normal operation.
10.
Move the sensor to its referenz position by pressing the
R key
.
NOTE
i
The sensor can always be moved back and forth between its front and center
positions with the R key.
11.
Move the sensor to its center position by pressing the
R key
.
12.
Measure distances
L1
and
L2
with a slide gage and calculate the remain-
ing offset as described on page 5-136.
– If the offset is still > ± 0.5 mm, the offset value must be corrected again.
Repeat the service routine from item 2.
– If the offset is < ± 0.5 mm, the calibration is OK.
z
Continue with service routine S034.5.
Quitting the service routine
XG
Plus
:
To return to the Service menu, press the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
XG 5:
To return to the Service menu, press the
up arrow key
above selection
field 3.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
T
T
R
R
R
R
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...