6.20 Replacing circuit boards
6.20
Case I:
X-ray tube assembly from an
external device,
ORTHOPHOS XG
Plus
System software status
V02.22 or higher.
CAUTION
– The replacement of an X-ray tube assembly from an external device is
possible if the X-ray tube assembly and the system have at least the
software version V02.22.
– The replacement is only possible within the same system class, i.e.
the X-ray tube assembly must come from a XG
Plus
system if it is to be
installed in another XG
Plus
system
1.
Switch the unit ON.
2.
Perform a software update to a software status
≥
V02.25.
3.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
Error message
E6 15 05
(undefined system serial number) is displayed.
4.
Acknowledge the error message with the
R key
.
Error message
E6 15 04
(undefined activation data) is displayed.
5.
Acknowledge the error message with the
R key
.
6.
Open the Service menu (see page 5-10).
7.
Start service routine
S008.3
(see page 5-50).
8.
Get the unit serial number from the rating plate of the unit and enter it (see
page 5-50).
NOTE
i
Any serial number which is unknown to the unit will not be accepted by the
unit. The serial number entered must be identical with the one on the rating
plate of the unit. If an inadmissible serial number is entered, the input will not
be accepted and the serial number can be entered again.
9.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
10.
Perform the complete system adjustment (see chapter 4):
– PAN adjustment
– For TSA units: TSA adjustment
– CEPH adjustment if cephalometer is installed
11.
Perform an acceptance test (applies to Germany only);
without consulting an expert
12.
Open the "Extended Details“ via SIXABCON.
This generates an XML file (with the system parameters) which is saved
under the network name of the system in the PDATA/P2K_Config folder
(see also section 1.10 on page 1-32).
z
The process is completed.
E6 15 05
(
R
R
E6 15 04
(
R
R
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...