båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
6 – 83
Tab 6
6.20 Replacing circuit boards
6.20
Case H: New X-ray tube assembly,
ORTHOPHOS XG
Plus
System software status
V02.22 or higher.
1.
Switch the unit ON.
2.
If the software status of the newly installed DX6 board is not compatible
with the current overall system software version, perform a software up-
date to the current system version as described in section 1.9.3.
3.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
Error message
E6 11 07
(undefined system class) is displayed.
4.
Acknowledge the error message with the
R key
.
The access level for the service menu is automatically started.
NOTE
i
Acknowledge any additional error messages with the
R key
.
5.
Press and hold down the Service key until the patient symbol keys light up
(approx. 2 s).
6.
Then press the patient symbol keys in the following order within 4 s:
b
–
d
–
a
.
Once the key combination has been entered correctly, service routine
017, test step 1 (select/confirm system class) is started automatically.
automatically. The
Memory button
is lit.
7.
Confirm the “Comfort“ system class (01):
To do this, first press the
Memory button
(R key lights up) and then
the
R key
.
8.
Quit the service routine via the
double arrow key
.
9.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
Error message
E6 15 05
(undefined system serial number) is displayed.
10.
Acknowledge the error message with the
R key
.
E6 11 07
(
R
R
R
a
b
d
Le
v
e
l 4
c
o
(
n
(
approx. 2 s
?
1
2
3
+
-
01
R
)
n
)
o
R
E6 15 05
(
R
R
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...