59 38 399 D3352
5 – 26
D3352.076.01.13.02
07.2008
5.5 Service routine S002
Tabs 5
S002.3: Radiation w/o rot. movement, max. radiation time selectable, prim. diaphr. opened fully
5.5.2
S002: Test step 3
(not for XG 3 / 3 PPE)
Difference from S002, test step 1:
9
Primary diaphragm is opened fully
NOTE
i
For selection and operating sequence, see S002, test step 1.
After you have selected test step 3, the primary diaphragm is moved to
the new position (opened fully).
During this time, a progress indicator is displayed in selection field 1.
n
(
1
2
3
1
2
3
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
3
S2
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...