båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 37
Tab 2
2.5 List of error messages
2.5
Location 42: Remote control, board DX42
Error code
Description
Actions required
see page
E6
42
01
General module initialization error
Note:
Error generated during module self-
test
z
Check unit configuration (with or without DX41)
via service routine S017.9,
configure correctly if necessary
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Replace board DX42
E6
42
02
Invalid system data or uninitialized
module storage data
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Replace board DX42
E6
42
03
Invalid commanding or control data
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Check software version of DX42 (in comparison
to overall software version) via service routine
S008.2,
perform a software update if necessary
z
Checking the CAN bus
z
Check the signal path from board DX1 to board
DX42,
replace module DX42 if necessary
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
42
04
Data transfer error or dialog error to
module (master side)
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
42
05
Data transfer error or dialog error to
bootloader of module
Note:
Occurs only in connection with
software update
z
Repeat software update
z
Checking the CAN bus
z
Replace board DX42
E6
42
06
Module failed in TTP*
(detected on master side)
z
Checking the CAN bus
z
Check the signal path from board DX1 to board
DX42,
replace module, if necessary
z
Replace board DX42
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...