3.9 Error analysis of X-RAY control signal path: unit HW version BA or higher
3.9
H321
is triggered at start of exposure.
3.9.2
Error messages
without installed remote
control
E3 13 40
occurs individually after unit is
switched ON.
Door contact
open?
• Close door contact
• Start exposure again
Is the error message displayed
again?
yes
yes
no
• Check door contact, repair if necessary
• Replace cable L117 (see section 6.21)
yes
no
Was the release but-
ton actuated during
switch ON?
Switch the unit OFF
and then ON again.
Make sure that the
release button is not
actuated during
switch-ON.
Do the error messages
recur?
yes
Short circuit in coiled cable of release button:
• Replace release button A2 (unit hardware
version
≥
BA)
Short circuit in cable L108:
Replace cable L108, (see section 6.21)
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...