59 38 399 D3352
2 – 34
D3352.076.01.13.02
07.2008
2.5 List of error messages
Tab 2
Location 41: Media interface card
Board DX41 is omitted as of unit hardware version BA
(see page 1-34).
E6
15
10
Update file for module is corrupt
z
Obtain current update file from the Sirona CSC
and perform software update via the SIRONA
home page
Error code
Description
Actions required
see page
E6
41
01
General module initialization error
z
If the error is a software error known to the Siro-
na Customer Service Center, a software update
(bugfix) must be performed
z
Replace board DX41
E6
41
02
Invalid system data or uninitialized
module storage data
z
If the error is a software error known to the Siro-
na Customer Service Center, a software update
(bugfix) must be performed
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Replace board DX41
E6
41
03
Invalid commanding or control data
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
41
04
Data transfer error or dialog error to
module (master side)
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
41
05
Data transfer error or dialog error to
bootloader of module
Note:
Occurs only in connection with
software update
z
Repeat software update
z
Checking the CAN bus
z
Replace board DX41
E6
41
06
Module failed in TTP*
(detected on master side)
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Replace board DX41
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...