4 – 47
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
Evaluating the X-ray image
4.
Evaluate the X-ray image:
Adjustment: not ok (length measurement with SIDEXIS)
S2
C
A2 = 44.3 ± 0.5 mm
A1 = 88.6 mm ± 1 mm
Adjustment: ok (length measurement with SIDEXIS)
± 0.75 mm
ZOOM: 1:1
A
B
A2 = 44.3 ± 0.5 mm
A2
A2
± 0.75 mm