båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 17
Tab 2
2.5 List of error messages
2.5
Location 11: Power PC, board DX11
Error code
Description
Actions required
see page
E6
11
01
Program sequence error
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and
perform such an update if necessary.
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Reset the entire unit adjustment (Pan, Ceph,
TSA) and readjust the unit
z
Replace board DX11
E6
11
02
Watchdog error
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Replace board DX11
E6
11
03
Operating system/resource error
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Replace board DX11
E7
11
04
Implausible data in EEPROM
z
Check the device configuration via service rou-
tines S017 and S018 and reconfigure if neces-
sary
z
Check adjustment with diaphragm test expo-
sures
If the adjustment is not OK...
z
Readjust the unit
If the adjustment is OK...
z
perform the individual system settings again
(e.g. programming of the patient symbol keys;
see operating instructions)
E6
11
05
RAM allocation failed
z
Replace board DX11
E7
11
07
Unknown or invalid definition of system
class
Note:
Occurs during first power-on after
replacement of board DX6 or DX11.
z
Perform actions required according to Chapter
"Replacing circuit boards".
E7
11
08
The user interface connected to this
unit does is not compliant with the
unit's system class configuration
z
Install the user interface compliant with the
system class configuration of this unit
E5
11
09
Internal error in program flow of board
DX11
z
Acknowledge error
If the error occurs again...
z
Perform a software update (bug fix)
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...