59 38 399 D3352
6 – 8
D3352.076.01.13.02
07.2008
6.1 Replacing the height adjustment motor (M1_4)/spindle
Tabs 6
Removing the remaining covers
z
Remove the top cover and pull connector X607 off of board DX1.
z
Now remove the following cover parts:
– Arm cover, top
– Slide cover, top rear
– Slide cover, bottom rear
– Slide cover, front
Attention: FH laser cable (PAN)!
NOTE
i
To prevent scratches, also remove the covers of the slide table and the drawer.