59 38 399 D3352
5 – 118
D3352.076.01.13.02
07.2008
5.15 Service routine S020 (XG
Plus
only)
Tabs 5
S020.1: Service for temple support
5.15
Service routine S020
(XG
Plus
only)
Service for temple support
This service routine can be used to change the factory-set switching points of
the temple width measurement used for setting the jaw width (“small”,
“normal” or “large”) within certain limits.
The following drawing shows the motor position values (initial and individual
offset ranges) as well as the mechanical end stops of the temple support.
Headrest version 1
(up to unit hardware version AG)
Headrest version 2
(as of unit hardware version BA)
The switching point between large and normal is referred to as the lower
switching point (marked L = lower) and the switching point between normal
and small as the upper switching point (marked U = upper).
These switching points can be set via service routine S20.1 within the follow-
ing limits:
Headrest version 1 (up to unit hardware version
≤
AG):
Lower switching point (L):
Setting range -10...+5
Upper switching point (U):
Setting range -5...+10
Headrest version 2 (up to unit hardware version
≥
BA):
Lower switching point (L):
Setting range -10...+5
Upper switching point (U):
Setting range -5...+13
NOTE
i
A negative offset shifts the respective switching threshold toward larger skulls;
a positive offset has the opposite effect.
Function S020
Service for temple support
Test step 1
Change switching thresholds of the temple support
0
34
46
60
large
-10 (24)
+5 (39)
normal
small
-5 (41)
+10 (56)
U
L
0
31
43
60
large
-10 (21)
+5 (36)
normal
small
-5 (38)
+13 (56)
U
L
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...