4 – 15
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
Coarse or precision adjustment?
Sensor adjustment can usually be performed directly via precision adjust-
ment. Only in exceptional cases, e.g. if one or several needles are completely
outside of the image field
C
in an image acquired with the
PRECISION
ADJUSTMENT
presetting, is it necessary to perform a coarse adjustment prior
to precision adjustment
D
. To do this, deactivate the
PRECISION
ADJUST
-
MENT
check box (see page 4-3) and then perform a coarse adjustment pro-
ceeding in the same way as for precision adjustment. The only difference
between coarse and precision adjustment is the size of the image area con-
sidered. Furthermore, there are fewer auxiliary lines in the coarse adjustment
mode.
X-ray image with unadjusted pan sensor:
Alternatively with precision adjustment setting (C) and coarse adjustment setting (D)
On the X-ray image with coarse adjustment (D), the center pin is just barely
visible on the left margin of the image field in the middle. Even in this extreme
case, an adjustment would still be possible.
NOTE
i
With SIDEXIS SW V01.45 and higher, a message window indicates whether a
coarse adjustment is required on completion of the precision exposure.
C
D
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...