59 38 399 D3352
6 – 62
D3352.076.01.13.02
07.2008
6.20 Replacing circuit boards
Tabs 6
DX6
(Tube assembly)
z
Replacing an X-ray tube assembly incl. DX6
z
Inserting a
new X-ray tube assembly
z
ORTHOPHOS
XG
Plus
z
System software version
V02.20
z
Please proceed as described in sec-
tion 6.20.3,
Case G
.
z
Inserting a
new X-ray tube assembly
z
ORTHOPHOS
XG
Plus
z
System software version
V02.22 or higher
z
Please proceed as described in sec-
tion 6.20.3,
Case H
.
z
Inserting an
X-ray tube assembly from an-
other unit
with software version V02.22
z
ORTHOPHOS
XG
Plus
z
System software version
V02.22 or higher
z
Please proceed as described in sec-
tion 6.20.3,
Case I
.
z
Inserting a
new X-ray tube assembly
z
ORTHOPHOS
XG 5 / 3 / 3 PPE
z
System software version
V02.20 or higher
z
Please proceed as described in sec-
tion 6.20.3,
Case J
.
z
Inserting an
X-ray tube assembly from an-
other unit
with software version V02.20
z
ORTHOPHOS
XG 5/3
z
System software version
V02.20 or higher
CAUTION
Never use a module from another unit in an XG 3
PPE unit or vice versa.
z
Please proceed as described in sec-
tion 6.20.3,
Case K
.
DX61
(Diaphragm unit)
z
Inserting a new DX61
z
Check the compatibility between the
software status of the modules and
that of the overall system. Perform
an update or a downgrade if neces-
sary.
NOTE
i
Please note that the specific diaphragm
unit required depends on the unit serial
numbers specified in the spare parts list.
DX7
(Easypad)
z
Inserting a new Easypad incl. DX7
z
ORTHOPHOS
XG
Plus
z
System software status
V02.20 or higher
z
It is
essential
that you perform a
software update to a system version
≥
V02.25
(see section 1.9).
NOTE
i
After the replacement of the Easypad,
the language settings on circuit board
DX7 are at the factory default
(00 = German, English, French,
Italian). If the configured system lan-
guage set (displayable via service rou-
tine S017.5 or the “extended detail que-
ry“ in SIXABCON) has a configuration
other than 00, this configuration will be
copied to board DX7 by the update func-
tion.
Board
Constellation
Measures
Page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...