59 38 399 D3352
6 – 80
D3352.076.01.13.02
07.2008
6.20 Replacing circuit boards
Tabs 6
6.20.3
Replacing an X-ray tube assembly incl. DX6
Case G: New X-ray tube assembly,
ORTHOPHOS XG
Plus
System software status
V02.20
1.
Switch the unit ON.
CAUTION
Do not acknowledge any error messages displayed at this point!
2.
Perform a software downgrade (via the main version) of the X-ray tube
assembly or board DX6 to software status V02.20 as described in section
1.9.3.
CAUTION
At this place do not perform
any software update
to a higher version
!
3.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
Error message
E6 11 07
(undefined system class) is displayed.
4.
Acknowledge the error message with the
R key
.
The access level for the service menu is automatically started.
NOTE
i
Acknowledge any additional error messages with the
R key
.
5.
Press and hold down the Service key until the patient symbol keys light up
(approx. 2 s).
6.
Then press the patient symbol keys in the following order within 4 s:
b
–
d
–
a
.
Once the key combination has been entered correctly, service routine
017, test step 1 (select/confirm system class) is started automatically.
automatically. The
Memory button
is lit.
7.
Confirm the “Comfort“ system class (01):
To do this, first press the
Memory button
(R key lights up) and then
the
R key
.
8.
Quit the service routine via the
double arrow key
.
E6 11 07
(
R
R
R
a
b
d
Le
v
el
4
c
o
(
n
(
approx. 2 s
?
1
2
3
+
-
01
R
)
n
)
o
R
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...