2.5 List of error messages
Tab 2
*)
TTP = Time Trigger Protocol
E6
81
73
Ceph
Sensor in ceph slot active on CAN
bus, but no hardware detection
z
Check cable L35/L38,
replace if necessary
z
Check plug X500 on board DX1,
replace board DX1 if necessary
z
Check Ceph sensor holder,
replace if necessary
E5
81
74
Ceph
Sensor in ceph slot not detected.
z
Check cables L35 and L38,
replace if necessary
z
Checking the CAN bus
z
Check connector X503 on board DX1,
replace board DX1 if necessary
z
Check ceph sensor holder, replace if necessary
E5
81
75
Ceph
Error while synchronizing image data
z
Perform a software update
If the error occurs again...
z
Check ceph sensor, replace if necessary
E5
81
77
Ceph
Voltage error detected on sensor.
(CCD voltages on board DX81P)
Note:
If this error occurs in connection with
other causal error messages (e.g. E6
6160, E6 9150), take appropriate
action
z
Check cable L35/L38,
replace if necessary
z
Check plug X503 on board DX1,
replace board DX1 if necessary
z
Check ceph sensor holder, replace if necessary
z
Perform a software update
z
Check ceph sensor, replace if necessary
E5
81
78
Ceph
Invalid data at start of exposure.
z
Acknowledge error and repeat procedure
If the error occurs again...
z
Perform a software update
z
Check version compatibility of XG software and
Sidexis,
perform a software update if necessary
z
Contact the Sirona Customer Service Center for
the settings/sensor information in the Xab.ini file
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...