5.17 Service routine S032
5.17
S032.10: Test function for sensor in PAN slot
Selecting the synchronized readout
sequence
2.
Select the mode of the desired synchronized readout sequence using the
arrow keys
of selection field 1.
HSIPan3x3 = sensor in the PAN mode with synchronized readout
sequence 3x3 clocked out
HSIPan4x4 = sensor in the PAN mode with synchronized readout
sequence 4x4 clocked out
HSITSA4x4 = sensor in the TSA mode with synchronized readout
sequence 4x4 clocked out
After the mode for the synchronized readout sequence has been selected the
Test key
(XG
Plus
) or the LED above the
Test key
lights up
(XG 5 / 3 / 3 PPE).
NOTE
i
If the R key (XG
Plus
) or the LED above the R key (XG 5 / 3 / 3 PPE) initially
lights up, the sensor holder must be moved to the corresponding position first.
To do this, press the
R key
.
Starting synchronized readout
sequence
NOTE
i
The sensor must be plugged in for at least 15 seconds before starting the
synchronized readout sequence.
3.
Start the synchronized readout sequence with the
Test key
.
CAUTION
The sensor must not be removed while the synchronized readout sequence is
active! Otherwise the electronics may be damaged.
1
2
3
+
-
HSIPan3x3
n
(
1
2
3
HSIPAN3X3______
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
(
n
(
n
T
10
T
T
R
R
T
T
1
2
3
Take Image...
n
(
1
2
3
TAKE_IMAGE_____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
10
S32
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...