59 38 399 D3352
D3352.076.01.13.02
07.2008
3 – 5
Tab 3
3.1 Error logging memory
3.1
3.1.1
Example of error logging data
System time
2006-03-06, 20:13:02
:
System time (clock on DX11)
Entry type
[Message]
:
General system event
[Error]
:
Error event
[Error Sidexis]
:
Network error event
[Stringname]
:
Free status texts
[Stringsegment]
:
Additional data (string names)
Entry data [Message]
Recording started
:
Beginning of a recording
Value: 9000
:
Sequence ID of recording
Recording stopped
:
End of a recording
Recording cancelled
:
Recording cancellation
Logbook started
:
Corresponds to power-on of unit
Image state switched to Released
:
Recording has been delivered to
and confirmed by SIDEXIS
Other entry data which document the occurrence of a rescue event include:
– Image state switched to Rescue
– Rescue request Sidexis Error
– Rescue request Sidexis TrackEpilogue
– Rescue request Sidexis Timeout
These entry data may also occur after "Recording stopped" or "Cancel" and
indicate an exceptional circumstance. You can supply important information
for error diagnosis in coordination with the Sirona Customer Service Center.
Entry data [Error]
E6 07 06
:
Error code
ERR_DX7_TTP_LOST
:
Cleartext display of error
Entry data [Error Sidexis]
SidErr: ERR_SOCKET_ERROR
:
Details on
network error
(for Sirona only)
SockErr:
:
Details on
network error
(for Sirona only)
Entry data [Stringname]
Key Act
:
Activation transaction
Key Ok
:
Activation transaction
Counter (XG 3 PPE)
:
Current counter reading
Entry data [Stringsegment]
7YFWDUFV-
:
e.g. activation or
E4MMRJBW
confirmation code
(for activation transaction)
061-00133
:
e.g. counter (ID counter reading)
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...