båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 17
Tab 5
5.2 Selecting a service routine
5.
2
5.2.1
Service routines with security access
NOTE
i
A security code is required for accessing service routines involving functions
such as radiation release or editing of configuration data or stored values.
This procedure prevents the inadvertent selection or activation of these
service routines.
To select a service routine or test step with security access, proceed as fol-
lows:
WARNING
Be sure to observe the radiation protection regulations applicable in
your country.
Selecting the service routine/test step
1.
Select the service routine or the test step and confirm your selection with
the
Service key
.
After you have confirmed your selection, a 0 appears in selection
field 2.
Confirming the security access
2.
Confirm the security access by once again selecting the number of the
main routine (in our example 2) with the
arrow keys
of selection field 2
and then pressing the Service key.
Following this double selection and confirmation, the service routine is ac-
tivated.
1
2
3
+
-
S002
1
n
(
1
2
3
S002_______1____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
(
o
(
o
+
-
1
2
3
2
n
(
1
2
3
___________2____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
(
o
(
o
+
-
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...