båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 27
Tab 2
2.5 List of error messages
2.5
E3
13
41
Release signal not applied on DX11
z
Check signal path for interruption according to
wiring diagrams, replace component if neces-
sary
z
Deactivate remote control with service routine
S017.6; connect release button directly to DX41
(instead of cable L17) and check for proper
functioning
z
Check release button
If the release button is functioning...
z
Check cable L17,
replace if necessary
If the release button is not functioning...
z
Check board DX41,
replace if necessary
E6
13
42
The hardware signal for radiation
release is applied on board DX1 dur-
ing unit operation even when no actu-
ated X-RAY release button is reported
via the CAN bus.
z
Check the X-ray signal path
E5
13 43
The door was opened during the
exposure.
z
Check the X-ray signal path
E5
13
50
Forehead support; has not left home
position despite movement pulses
z
Check light barriers V5_1 to V5_4,
replace headrest or
light barrier if necessary
z
Check cable L18,
replace if necessary
z
Check board DX5,
replace if necessary
E4
13
51
Forehead support; has not left home
position and no motor movement
z
Check motor M5_1,
replace headrest if necessary
z
Check light barriers V5_1 to V5_4,
replace headrest or
light barrier if necessary
z
Check cable L18,
replace if necessary
z
Check board DX5,
replace if necessary
E5
13
52
Forehead support; has left home posi-
tion without detected movement
pulses
z
Check light barriers V5_1 to V5_4,
replace headrest or
light barrier if necessary
z
Check cable L18,
replace if necessary
z
Check board DX5,
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...