6.20 Replacing circuit boards
Tabs 6
6.20.1
Measures following replacement of boards
CAUTION
After replacing boards or modules containing boards, check to make sure that
the software version of the module corresponds to the current software status
of the system. The software version for the modules can...
for XG
Plus
:
be accessed through service routine S008.2 or the extended detail query in
SIXABCON. You can also check the info screen in advance to determine
whether the current software constellation is permissible. If this is not the
case, the version number of the entire software is labeled with an asterisk (e.g.
V02.20*)
for XG 5 / 3 / 3 PPE:
be accessed through service routine S008.2 or the extended detail query in
SIXABCON.
Perform a software update or downgrade (see Section 1.9) in case of software
incompatibilities.
Always perform the measures described below in the given sequence and do
not carry out any other actions between the steps.
DX32
DX61*
DX11
DX1
DX91
DX41**
DX42
see page 5-9
DX71
*
not with XG 3 / 3 PPE
** from system hardware version BA
does not apply to circuit board DX41
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...