båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 33
Tab 2
2.5 List of error messages
2.5
Location 15: Configuration, update
Error code
Description
Actions required
see page
E7
15
01
Wrong memory modules.
If a DRAM memory module is plugged into board
DX11...
z
Replace memory module or DX11
If no DRAM memory module is plugged into board
DX11...
z
Replace DX11
E7
15
02
Wrong module constellation for this
unit.
Incompatible modules were used.
z
Refer to list of software versions
z
Check to make sure the correct modules are in-
stalled (e.g. after replacing modules)
z
Perform a software update
If the error occurs again...
z
Contact the Sirona Customer Service Center
(CSC)
E7
15
03
Wrong software constellation of mod-
ules.
z
Check software versions of unit (info screen or
service routine S008.2) and of SIDEXIS XG and
Perform or repeat software update or down-
grade if necessary
E6
15
04
Product activation keys invalid or not
available.
Note:
Occurs after replacement of tube
assembly (DX6) or DX11 and possibly
after software updates.
See also chapter "Measures following
replacement of boards" starting on
page 6-60.
z
Execute service routine S008.4 and confirm the
existing activation profile of the unit.
E6
15
05
Unit serial number invalid or not avail-
able.
Note:
Occurs during first power-on after
replacement of board DX6 or DX11.
See also chapter "Measures following
replacement of boards" starting on
page 6-60.
z
Execute service routine S008.3 and confirm or
enter the unit serial number at the unit.
E6
15
06
Wrong or invalid manufacturer code
on an module was detected.
z
Replace board DX11
z
Replace the X-ray tube assembly
E6
15
07
The user interface or the remote con-
trol is not compatible with the unit.
HINWEIS
i
An Orthophos XG5 (Multipad) or XG-
Plus
(Easypad) as well as an Ortho-
phos remote control must be installed.
z
Install the correct user interface or remote
control.
If necessary, obtain a new user interface or re-
mote control from the manufacturer.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...