7 – 3
Tab 7
7
Maintenance
DANGER
Shock hazard! It is essential that you switch the X-ray unit OFF before
replacing any components!
The unit must be disconnected from the junction box of the building
installation before replacing any parts near the power supply, power
switch, board DX 32 or the X-ray tube assembly!
CAUTION
– Switch the X-ray unit OFF before connecting a measuring instrument.
– Select the correct current/voltage type and adjust the measuring range to
match the expected readings.
– Perform continuity tests only on units which are switched off.
– Observe the prescribed cool-down intervals if several exposures must be
taken to check a measurement.
CAUTION
Please observe the usual precautionary measures for
handling boards (ESD).
Touch a ground point to discharge static electricity
before touching any boards.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...