 pHTestr

®

 5 Pocket Testers       

Features

•  Large easy-to-read LCD with 

3-color backlight

•  Accuracy of ±0.01 pH ± 1 digit
•  1, 2, or 3 calibration points
•  Auto buffer recognition:  

US or NIST

•  Stability and battery status 

icons

•  Replaceable single-junction 

sensor saves you money

•  Auto power-off function  

conserves battery life

• IP67 waterproof rating

Operating Manual

 

Model pH5 — Glass bulb probe for general liquid testing 
Model pH5F — Flat probe for surface or micro-volume testing 
Model pH5S — Spear-tip probe for testing semisolid and  

  solid materials

Thank you for purchasing an Oakton

®

 pHTestr

®

 Pocket pH Tester. 

Please carefully read this operating manual before using the product 

to obtain an accurate and reliable test result and avoid unnecessary 

damage to the tester or  

sensor/probe.

the tester is not in use to maintain the sensor’s accuracy. To achieve 

maximum accuracy, we recommend soaking the probe in the 3M 

KCL storage solution overnight (12 hours) to activate the glass  

membrane thoroughly. 

Storing the pH5 and pH5F testers dry will NOT cause any permanent 

damage. It will only temporarily cause the probe to lose its sensitivity, 

which can always be restored by soaking in the storage solution.

Model pH5S:

 Never store the spear probe in a dry condition because 

permanent damage can be caused. The spear probe should always 

be stored in the 3M KCL soaking solution (potassium chloride).  

One 10-mL bottle of storage solution comes with the tester kit. If the 

soaking solution was contaminated, please replace immediately.

pH Calibration

1. Short press  /MEAS key to turn on. 

2.  Rinse the probe in distilled water and use tissue paper  

to gently dab off excess water. 

3.  Pour certain amount (about half volume of the calibration bottles) 

of pH 7.00 pH and pH 4.00 buffer solutions in the corresponding 

calibration bottles.

4.  Long press CAL/

  

key to enter calibration 

mode, short press  /MEAS key to return to 

measurement mode.

5.  Dip the probe in pH 7.00 buffer solution,  

stir gently, and allow it to stand still in the 

buffer solution until a stable reading is 

reached. When measurement stability icon 

(

) appears and remains on the screen  

(see Fig 1); then short press CAL/

  key to 

complete the one-point calibration. The  

tester returns to measurement mode and 

indication icon “

M

” will appear at the  

bottom left of the screen, indicating the  

middle point of calibration has been completed.

6.  For 2-point calibration, start with pH 7.00 buffer solution; once 

confirmed then rinse the probe in distilled water and dry it with 

tissue paper. Dip the probe into pH 4.00 buffer solution, follow the 

steps above to complete 2-point calibration. Indication icons “

L

” 

and “

M

” will appear on the bottom left of the screen. 

7.  If necessary to perform a 3-point calibration, rinse probe with 

water, dip the probe into pH 10.01 buffer solution (sold sepa-

rately), and follow the steps above. Indication icons “

L

”, “

M

”,  

and “

H

” will appear on the bottom left of the screen.

Notes

A.  Tester can perform 1 to 3 points automatic calibration. Please note 

that pH 7.00 (USA Standard) or pH 6.86 buffer solution (NIST Stan-

dard) must be used to conduct 1st point calibration. Then use other 

buffer solution to conduct 2nd or 3rd point calibration. Tester will 

automatically recognize five kinds of pH buffer solutions. For details, 

please refer to the following table:

Calibration

USA series

NIST series

Indication  

icons

Recommended

1-point

7.00 pH

6.86 pH

M

Accuracy: 

0.1 pH

2-point

7.00 pH, 

4.00 pH or 

1.68 pH

6.86 pH, 

4.01 pH or 

1.68 pH

L    M

Range: 

<7.00 pH

7.00 pH, 

10.01 pH or 

12.45 pH

6.86 pH,  

9.18 pH or 

12.45 pH

M    H

Range: 

<7.00 pH

3-point

7.00 pH, 

4.00 or 1.68 pH, 

10.01 or 

12.45 pH

6.86 pH,  

4.01 or  

1.68 pH, 9.18  

or 12.45 pH

L    M    H

Range:  

0 to 14.00 pH

B.  pH 4.00 and pH 7.00 buffer solutions are included in the pH5S test 

kit only but pH 10.01 is not. Users can purchase it separately if 

needed. The buffer solutions poured into the calibration bottles 

are NOT for one-time use. They can be used multiple times as 

long as they are not contaminated, and the bottles are covered 

when not in use. After that, we recommend replacing the buffer 

solutions in the calibration bottles with new ones that are in the 

buffer bottles (50 mL) to keep the accuracy of the standard buffer 

solutions. Do not pour used buffer solutions back into the buffer 

bottles in case of contamination.

C. The tester has self-diagnostic functions:

Symbol

Self-diagnostic information

Checking and how to fix

Er 1

Wrong pH buffer solution or 

the range of calibration  

solution exceeds standard.

1.  Check whether pH buffer  

solution is correct (1st point 

calibration must be 7.00).

2.  Check whether the probe is  

damaged.

3.  Check if there is any air 

bubble in the glass bulb 

sensor.

Er 2

Press CAL/

  key when 

reading is not stable during 

measurement.

Wait for the measurement 

stability icon (

) to appear and 

stay, then press CAL/

  key.

* If you find any air bubble in the glass bulb of the pH sensor, simply invert the 

probe for a few times to remove it. The existence of an air bubble in the glass 

bulb will significantly decrease the accuracy of measurement. 

*  The 1st point calibration must be 7.00 pH. Perform the 2nd point calibration 

(4.00 pH) immediately after the 1st point. Do NOT turn off the meter before 

you conduct 2nd point calibration. If the meter is turned off after 1st point  

calibration, user will need to restart the calibration process with the 7.00 pH 

first and the 4.00 pH following after. Calibrating directly in pH 4.00 after  

turning meter off and back on will cause “Er1”.

pH Measurement

1.  Short press  /MEAS key to turn on the tester. Rinse the probe in 

distilled water, and gently dab off excess water with tissue paper 

(do not rub or wipe the glass sensor). 

2.  Dip the probe in sample solution, stir gently, and allow it to stand 

until a stable reading is reached. Take readings after stability indi-

cator icon (

) appears and stays on the screen.

Notes

A. Applications of each model:

Model/Probe

Application

pH5/Bulb probe

Regular water solutions’ pH measurement such as 

hydroponics, aquaculture, pools and spas, water 

treatment, brewing, etc.

pH5F/Flat probe

Flat surface measurement such as skin, paper, fabric, 

leather and so on; micro sample testing;  

also works well for regular water solution.

pH5S/Spear-tip 

probe

Cheese, sushi rice, meat, fruit, bread, soil, solid  

culture medium and semisolid medium measurement;  

also works well for regular water solution.

B. 

Model pH5F Flat probe tester 

is mostly for flat surface sample 

testing.

•  For skin test: skin should be without sweat or dirt, nor be overly 

cleaned (do not use facewash products before testing) to avoid 

affecting measurement results. Dampen skin with some distilled 

water, slightly force flat probe onto the skin, get readings after 

value stabilized.

•  For paper, fabric and leather test: add 1 to 2 drops of distilled water 

on surface, then perform measurements.

•  For micro sample testing, use a container with an inner diameter 

19 mm and a flat bottom. The tester can test volume 

0.5 mL.

C. 

Model pH5S Spear-tip probe tester

 is widely used for solids  

containing water or semisolid medium. When conducting such tests, 

pay attention to insert probe evenly and be careful to prevent probe 

from damage. If the medium is too hard (such as meat or fruits), 

please bore a small hole with a knife before inserting the probe.  

For pH measurement of any food (such as meat, cheese, sushi, etc.), 

it should be a sampling test. That means test samples should no  

longer be edible.

Special Notes

•  The pH probe must be rinsed thoroughly after each use. Soap water 

should be used to clean off any grease or other contaminants.

•  These pH testers will 

NOT

 give accurate or stable pH readings 

when testing distilled or deionized water. This is because distilled 

or deionized water do not have enough ions present for the elec-

trode to function properly. To measure distilled or deionized water’s 

pH, users need to use a specialized instrument technique. When 

testing purified water like spring water or drinking water, it will take 

longer for the readings to get stabilized (typically 3 to 5 minutes) 

because there are very few ions left to be detected by the sensor in 

those purified water.

•  Do 

NOT

 store probe in distilled or deionized purified water because 

that will cause permanent damage to the pH probe. Purified water 

is only recommended for rinsing the probe. The probe should be 

stored in 3M KCL pH electrode storage solution (SKU 00653-04) for 

best accuracy.

Setting the Parameters

When tester is turned off, long press  /MEAS key to enter parame-

ter setting. Short press MODE/

 key to switch from P1 to P2…P7. 

Short press CAL/

  key and parameter will flash, then short press 

MODE/

 key to choose desired parameter. Short press CAL/

  key  

to confirm parameter selection. Long press  /MEAS key to return  

to measurement mode.

Symbol

Menu setting

Selection

Factory 

default

P1

Select pH buffer solution

USA – NIST

USA

P2

Set low measurement alarm

0 to 14.00 pH

0

P3

Set high measurement alarm

0 to 14.00 pH

14.00

P4

Select Automatic Lock (HOLD)

Off – On

Off

P5

Select backlight

Off – 1 – On

1

P6

Select temperature unit

°C – °F

°C

P7

Restore to factory default

No – Yes

No

Notes

A. Select standard pH buffer solution (P1)

There are two options of standard buffer solutions: USA series and 

NIST series. 

B. High and Low Measurement Heads-Up Examples (P1 and P2)

Alert when measured value 

3.20 pH:

 Preset lowest value (P2) = 

3.20 pH, highest value (P3) = 14.00 pH, when measured value is 

3.20 pH (stability icon (

) displays on LCD), the red backlight 

appears on display.

Alert when measured value 

8.60 pH:

 Preset highest value (P3) 

= 8.60 pH, lowest value (P2) = 0.00 pH, when measured value is 

8.60 pH (stability icon (

) displays on LCD), the red backlight 

appears on display.

Alert when measured value 

3.20 pH or 

8.60 pH:

 Preset lowest 

value (P2) = 3.20 pH, highest value (P3) = 8.60 pH, when mea-

sured value is lower than 3.20 pH or higher than 8.60 pH (stabil-

ity icon (

) displays on LCD), the red backlight appears on 

display.

C. Automatic Lock (P4)

Select “On” to activate auto lock function. When reading is stable for 

more than 10 seconds, the tester will lock the value automatically, 

and “

HOLD

” icon will appear on the bottom left of the screen. Press 

CAL/

  key to cancel HOLD on reading.

D. Backlight (P5)

Select “Off” to turn off backlight function, “On” to turn on backlight 

function, or “1” to have backlight last for 1 minute.

E. Temperature Unit (P6)

Select °C or °F; the factory default is °C.

F. Factory Default Setting (P7)

Select “Yes” to recover instrument calibration to the theoretical 

value (pH value in zero potential is 7.00 pH, slope is 100%), parame-

ter setting returns to initial value. This function can be used when 

instrument does not work properly in calibration or measurement. 

Calibrate and measure again after resetting the unit to factory default 

status.

ORP Measurement

ORP stands for Oxidation-Reduction Potential. ORP is a measure of 

the cleanliness of the water and its ability to break down contami-

nants. Note that ORP sensor module 35634-58 must be purchased 

separately (see 

Ordering Information

).

Once ORP sensor probe is attached to tester, press MODE/

 

key to 

enter ORP mode. Rinse the probe in distilled water and dry it. Dip 

the probe in sample solution, stir gently, and allow it to stand still 

until a stable reading is reached. Get readings after measurement 

stability icon (

) appears and remains on the screen.

6.6

7

pH

pH

Figure 1

Keypad Functions

Short press

 = <2 seconds         

Long press

 = >2 seconds

MEAS

1.  Short press to turn on the tester and long press to turn  

off the tester.

2.  When turned off, long press to enter parameter setting mode. 

3.  In measurement mode, short press to turn on backlight.

MODE

1.   In measurement mode, short press to switch parameter from 

pH

 to 

ORP

 (Oxidation Reduction Potential); ORP probe sold 

separately.

2.  In parameter setting mode, short press to change parameter 

(unidirectional).

CAL

<

1. Long press to enter calibration mode.

2. In calibration mode, short press to confirm calibration.

3.  When measured value is locked (HOLD icon), short press to 

unlock.

Preparation Before Use

Models pH5 and pH5F:

 If it is first-time use or the tester hasn’t been 

used for a long time, pour some 3M KCL solution to the Fill line in 

the probe cap and soak the probe for about 15 to 30 minutes. Users 

can store the probe in the 3M KCL solution in the probe cap when 

1