båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
7 – 27
Tab 7
7.8 Checking the diaphragm
7.
8
SIDEXIS screen
The exposure dialog box showing the exposure status appears in
SIDEXIS.
XG
Plus
:
Display on the Easypad
Service routine
S010.5
and its specific exposure parameters are
displayed on the Easypad touchscreen.
XG 5:
Display on the Multipad
The single-line display of the Multipad shows the exposure parameters
of service routine
S010.5
.
Starting the exposure
3.
Take an exposure (80 kV/14 mA; 0.60 s):
– Press the
R key
to move the unit back to the starting position.
XG
Plus
:
The procedure is completed when the status message “ORTHOPHOS is
ready for exposure“ appears on the touchscreen and the exposure
parameters of service routine
S010.5
are displayed.
XG5:
The procedure is completed when the exposure parameters of service
routine
S010.5
are displayed and the progress indicator disappears.
– Press the release button. Hold down the release button until image ac-
quisition is completed and the preview image appears in the exposure
dialog box.
?
1
2
3
+
-
80 kV/14 mA
0.60 s
S10
5
1
2
3
80/14___0.60____
R
WAIT
!
UNTIL
UNIT
IS
IN
STARTING
POSITION
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...