2.5 List of error messages
2.5
2.5
List of error messages
HINWEIS
i
In the following table, the error codes are sorted by the location or function unit
where the error has occurred. For enhanced clarity, the corresponding ID in
the error code is printed in bold type.
Location 06: Tube assembly / Board
Error code
Description
Actions required
see page
E6
06
01
General error during module initializa-
tion
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
If the error occurs frequently,
replace the X-ray tube assembly
E6
06
02
Invalid system data or uninitialized
module storage data
z
Perform service routines S005.2
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
06
03
Invalid commanding of control data,
CAN bus error
Note:
This error may also occur in connec-
tion with other causal error messages!
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
06
04
Data transfer error or dialog error to
module (master side)
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E6
06
05
Data transfer error or dialog error to
bootloader of module
Note:
Occurs only in connection with soft-
ware update
z
Repeat software update
z
Checking the CAN bus
z
If the error occurs repeatedly or the module is
no longer addressable, replace
the tube assembly
E6
06
06
Module failed in TTP*
(detected on master side)
Note:
This error may also occur in connec-
tion with other causal error messages.
Please also observe the causal error
message! It appears only after you
acknowledge the first error message.
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
z
If the error occurs repeatedly or the module is
no longer addressable, replace
the tube assembly
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...